User's Manual

Ultraview Care Network
7-6
Display Detail
The User Settings screens list all user-configurable options for ECG, respiration,
Varitrend 3, temperature, invasive pressure, non-invasive blood pressure, and
SpO
2
. From these screens you can change and store the options in your module.
Figure 7-1 illustrates the ECG User Settings screen which is used to change and
store ECG rate alarm limits. In this example, LEARNED has been chosen as the
adult rate alarms type.
The parameter configuration tables in this chapter list all available options, the
available user settings, and the factory defaults for each parameter.
Figure 7-1: ECG user settings
CURSOR AND CURSOR keys scroll through the two columns of
parameters on the screen.
and keys scroll the list of valid settings for the highlighted parameter.
PREV PAGE key returns to the previous screen.
NEXT PAGE key advances to the next screen.
PRINT PAGE key prints the current screen.
STORE key stores the currently selected settings as defaults.
PARAMETER DESCRIPTION VALUE PARAMETER DESCRIPTION VALUE
ECG RATE ALARMS (ADULT) ECG RATE ALARMS (INFANT)
TYPE LEARNED TYPE LEARNED
HIGH RATE 180 HIGH RATE 220
•LOW RATE 40 •LOW RATE 100
HIGH RATE OFFSET 50 HIGH RATE OFFSET 50
LOW RATE OFFSET 50 LOW RATE OFFSET 50
•DELAY 3
Set ECG alarm type to LEARNED or FIXED limits

STORE
SELECT PARAMETER
CURSOR CURSOR
CHANGE VALUE
PAGE X OF X
PRINT
PAGE
NEXT
PAGE
PREV
PAGE