User's Manual

Module Configuration Manager
7-29
Table 15: Pressure Parameter Configuration Table (Channel 4)
Parameter Available User Settings Factory Defaults
User
Settings
Label (P4)
I C P, L A P, R A P, CV P, PA , ART, U A ,
UV, or PRS
CVP
Scales On/Off/Auto Off
Top of Scale 10 to 500 mmHg 50 mmHg
Filter Frequency
3, 8, 12, 15, 18, 21, 25, 30, 35,
40 Hz
12 Hz
Sweep Speed
50, 25, 12.5, 6.25 mm/sec, or
same as ECG
Same as ECG
Artifact Rejection On/Off Off
Numeric Size
Systolic/Diastolic Large
Mean Large
All Large
Systolic/Diastolic
Large
Alarm Type Learned/Fixed Learned
Systolic Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 50 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 10 mmHg
Diastolic Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 20 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 0 mmHg
Mean Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 20 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 5 mmHg
CPP Alarm On/Off Off
High Alarm Limit -45 to 300 mmHg or Off 50 mmHg
Low Alarm Limit -50 to 295 mmHg or Off 0 mmHg