User's Manual

Ultraview Care Network
8-14
Alarms
Table 2 displays the conditions for which an alarm may be initiated.
Selecting ECG Leads
When you select a precordial lead, a help message appears that describes the
proper location for the chest electrode. For example, if you select V1, the
message (C) 4TH INTERCOSTAL SPACE, RIGHT STERNAL BORDER is
displayed.
No message appears when you choose to display a limb or augmented lead (for
example, I, II, III, AVR, AVL, or AVF).
When you change lead selections, the new waveform is displayed and the learn
sequence is automatically initiated. If you change lead selections while processing
is suspended, the module initiates the learn sequence when you resume
processing.
Table 2: ECG Alarms
Alarm Cause Of Alarm Display Indication
VFIB Ventricular fibrillation detected VFIB
Asystole >5-seconds with no QRS detected Heart rate 0; flashing low rate limit;
ASYSTOLE
High Heart Rate High limit violated Flashing high limit and
HI RATE ALARM
Low Heart Rate Low limit violated Flashing low limit and
LO RATE ALARM
Chan 1 & 2 - Leads Off One or more of the electrodes for
both ECG leads are off or loose
No ECG trace and
CHAN 1 & 2 LEADS OFF
Chan 1 - Lead Off One or more of the electrodes for the
first ECG lead are off or loose
No ECG trace and
CHAN 1 - LEADS OFF
Chan 2 - Lead Off One or more of the electrodes for the
second ECG lead are off or loose
No ECG trace and
CHAN 2 - LEADS OFF
Rate Source Unavailable Heart rate is not available from either
the primary or any alternate heart
rate source
HR UNAVAILABLE
Noisy Signal Excessive noise, artifact or baseline
shift
NOISY SIGNAL
Low ECG Voltage QRS amplitude is between 0.16 and
0.20 mV (adult only)
ECG VOLTAGE TOO LOW
Changing your selection for the first lead may change the
selection for the second lead.
To change the lead selection:
1 Touch ECG.
2 Touch LEAD CONTROL.
3 Touch 1ST LEAD or 2ND
LEAD.
4 Select lead.