User's Manual

ECG
8-17
Changing the Display Resolution
The MONITOR/EXTENDED key determines the display resolution of the ECG
traces.
The Monitor mode displays a reduced frequency range, but filters out more
noise for better viewing.
The Extended mode displays a broader frequency range, and also more
noise.
The factory default setting for display resolution is monitor mode.
Selecting Primary and Alternate Heart
Rate Source
You may select a primary rate source as well as enable alternate rate sources.
Rate sources include ECG, ART (arterial pressure), UA (umbilical artery) and
SpO
2
. If you enable one or more alternate sources, a heart rate will be displayed,
if available, from either the primary or an alternate source. If you do not enable
any alternate sources, a heart rate will only be displayed if the primary rate source
is available.
If a heart rate is not available from any source, the message HR UNAVAILABLE is
displayed in the ECG waveform zone, and ??? replace the rate numbers.
Changing the display resolution does not change the waveform
bandwidth used to analyze the ECG signals for arrhythmia and
ST segment level.
Table 3: Display Resolution
Key Telemetry Non-telemetry
MONITOR 0.5 — 30 Hz 0.5 — 40 Hz
EXTENDED 0.05 — 30 Hz 0.05 — 150 Hz
If ART is to be used as an alternate heart rate source,
Spacelabs Medical recommends setting up each monitor with
only one arterial pressure channel.
•Use of SpO
2
as the primary heart rate source is not
recommended by Spacelabs Medical.
CAUTION:
Cardiovascular Artifact (CVA) detection is disabled for the
respiration channel when using an alternate heart rate
source.
To change the display resolution:
1 Touch ECG.
2 Touch SETUP.
3 Select MONITOR or
EXTENDED.
To select a primary heart rate
source:
1 Touch ECG.
2 Touch SETUP.
3 Touch RATE SOURCE.
4 Select the desired primary rate
source (ECG, ART, UA, or
SPO2).
To enable alternate rate sources:
1 Touch ECG.
2 Touch SETUP.
3 Touch RATE SOURCE.
4 Set the desired alternate
source(s) to ON (ECG, ART,
UA, and/or SPO2).