User's Manual

Ultraview Care Network
8-18
Suspending/Resuming ECG Processing
When you touch YES in the Suspend Processing menu, ECG and respiration
waveforms continue to be displayed, but no processing occurs. If you touch NO,
the display returns to the ECG menu without affecting ECG and respiration
processing.
When you suspend ECG processing:
the message ECG PROCESSING SUSPENDED appears in the ECG
waveform zone. In modules with the respiration option RESP PROCESSING
SUSPENDED appears in the respiration waveform zone.
question marks replace the heart rate immediately and replace the ST
segment value after 30-seconds. If an alternate heart rate source is available,
the new heart rate is displayed. Question marks (???) will continue to display
for the ST segment value.
the message ECG ALM OFF replaces the alarm limits (if no alternate rate
source is available).
the keys in the ECG Alarm Limits menu display dotted outlines to indicate
they are disabled (if no alternate rate source is available).
the key SUSPEND PROCESSING changes to RESUME PROCESSING.
Printing ECG Recordings
All printouts of ECG data are annotated with the following:
Bed identification
Time and date of the printout
Lead designator
Depending on the printer model and software version, the ALL LEADS recording
can be printed on a single page or as consecutive two- or four-channel recordings.
For these recording formats, 6.25-seconds of waveform data will be printed for all
available leads.
If you are using a 90838 PrintMaster and if all the leads are connected, touching
the STAT REPORT key will provide a 12-lead report. For this recording format,
2.5-seconds of waveform data will be printed for each lead plus 10-seconds of
waveform data for a rhythm lead.
A 12-lead report can be scheduled to print at regular time intervals using the
AUTO REPORT and AUTO PRINT features. Available intervals are 30-minutes,
1-hour, 2-hours, 4-hours, 8-hours, and 24-hours.
A printer (with 1.10.04 version software or greater) is required
for the STAT REPORT feature.
A 12-lead report cannot be printed using the 90449 printer
module.
To suspend ECG processing:
1 Touch ECG.
2 Touch SUSPEND
PROCESSING.
3 Touch YES.
To resume ECG processing:
1 Touch ECG.
2 Touch RESUME
PROCESSING.
3 Touch YES.
To print recordings of all leads:
1 Touch ECG.
2 Touch PRINT.
3 Touch ALL LEADS.
To print a 12-lead report:
1 Touch ECG.
2 Touch REVIEW.
3 Touch REPORT REVIEW.
4 Touch STAT REPORT.
5 Touch PRINT.
To print a 12-lead report at
pre-selected intervals:
1 Touch ECG.
2 Touch REVIEW.
3 Touch REPORT REVIEW.
4 Select AUTO REPORT of
30-minutes, 1-hour, 2-hours,
4-hours, 8-hours, or 24-hours.
5 Touch AUTO PRINT ON.