User's Manual

Ultraview Care Network
2-6
Controlling Alarms via the
TONE RESET/ALM SUSPEND Key
During an Alarm Condition
If you touch TONE RESET/ALM SUSPEND once at a monitor during an alarm
condition:
The alarm tone is silenced at that monitor for 45-seconds, but the alarm tone
continues to sound at other monitors. If the alarm condition continues, the
alarm tone resumes at that monitor at the end of the 45-second period. If
another parameter goes into alarm during this period, or if the initial condition
ends and then begins again, the alarm tone will again sound at that monitor.
The message ALARM TONES SUSPENDED FOR 45-SECONDS is
displayed only on that monitor.
All flashing parameter keys continue to flash.
At a bedside monitor, when you touch the TONE RESET/ALM SUSPEND again
within 45-seconds:
All alarms and alarm recordings are suspended for 3-minutes. No alarms
sound for any reason during this period, and no new alarm recordings are
produced.
A message is displayed to indicate that all alarms are suspended.
Trend data may not be collected. This feature can be enabled or disabled
only by your system administrator.
Normal alarm operation can be resumed within the 3-minute time period by
touching the TONE RESET/ALM SUSPEND or RESUME ALARMS key again.
In the Absence of an Alarm Condition
Touching the TONE RESET/ALM SUSPEND key once when no parameters are in
an alarm condition suspends the alarms at the bedside monitor. (Refer to During
an Alarm Condition on page 2-6.)
The ALM SUSPEND feature is only available at bedside
monitors. If this feature has been disabled (set to OFF), the
TONE RESET/ALM SUSPEND key does not suspend alarms.
The ALM SUSPEND key is relabeled as RESUME ALARMS
after the first key touch.
The TONE RESET/ALM SUSPEND key is relabeled as ALARM
SUSPEND after the first key touch.
To silence an alarm tone for
45-seconds at any monitor, touch
the TONE RESET/ALM SUSPEND
key once.
To resume alarm tones at a
central monitor, touch the TONE
RESET/ALM SUSPEND key again
.
To suspend all alarms at a
bedside monitor for 3-minutes,
touch the TONE RESET/ALM
SUSPEND key again within
45-seconds.
Note: If alarms resume the
second time the key is touched,
your bedside monitor may be
configured differently for this
feature. Contact your system
administrator for details.
To resume normal alarm
operation at a bedside monitor
within the 3-minute period, touch
the TONE RESET/ALM SUSPEND
or RESUME ALARMS (Ultraview
1030, 1050, 1500, and 1600
monitors) key again.
To suspend alarms at a bedside
monitor when no alarm is
sounding, touch the TONE
RESET/ALM SUSPEND key once.