User's Manual

Ultraview Care Network
2-10
Automatic Recording of an Alarm
Touch the ALARM PARAMS key to display a menu of parameters that can initiate
alarm recordings on a bedside recorder and on a network printer if the parameter
goes into an alarm condition. Recording on a network printer is configuration-
dependent. See your system administrator for details.
If alarm recordings are disabled for a parameter, no alarm recording is produced
on either the bedside or network printer. However, the designated alarm tones are
still generated and alarm messages are displayed.
Alarm recordings print the last 12- or 20-seconds (selectable under RECORDING
DURATION) of a waveform or print until the alarm stops or is suspended,
whichever is longer. To stop a recording at the bedside printer, touch STOP or
UNIT OFF on the printer module.
When a bedside recorder is present in the monitor and is selected and online, an
alarm recording is automatically directed to that recorder.
Use the key labeled OTHER ON/OFF for any new parameter not yet on the alarm
parameters menu.
Spacelabs Medical strongly recommends verifying alarm
parameter settings on the monitor once per shift and upon each
admission.
If a printer is not attached to the monitor, the recorder channel
(1, 2, 4) keys do not appear in the menu.
To select parameters for alarm
recording and generate
automatic recordings on a
bedside recorder:
1 Touch MONITOR SETUP.
2 Touch RECORDER CONFIG.
3 Select RECORDING
DESTINATION.
4 Select a destination for the
alarm recording.
5 Touch PREVIOUS MENU.
6 Touch ALARM PARAMS.
7 Select parameter key(s) ON to
initiate a recording in case of
alarm. If the parameter key is
OFF, no alarm recording is
produced for that parameter in
an alarm condition.