Alarms Troubleshooting Guide
Clinical Situation Possible Cause Solution
Duplicate alarm recordings
print at the bedside and system
printer
Alarm recording directed to both
bedside and network printers
under the RECORDING
DESTINATION key.
Select THIS MONITOR or
NETWORK for alarm recordings.
No alarm recordings are
printed
Parameters for alarm recordings
are set to OFF.
Select parameters for alarm
recordings from the Alarm
Parameters menu or Record
Alarm Parameters menu.
Bedside printer is OFF.
The alarms have not been directed
to the bedside printer.
ALARMS SUSPENDED for
3-minutes.
Turn bedside printer ON.
Ensure that THIS MONITOR or
BOTH is selected.
Ensure that alarms are not
suspended.
System printer is not identified as
one of two network printers.
Have your system administrator
verify configuration.
Printer is out of paper. Load paper into printer.
Alarms continue to violate after
touching TONE RESET
More than one alarm may be in
violation.
Touch TONE RESET/ALM
SUSPEND again.
No alarm notification occurs at
a central monitor
The Alarm Watch function is not
active.
Activate Alarm Watch for the
desired beds at the central.
Parameters not displayed. Display the desired parameters
using SCREEN FORMAT under
MONITOR SETUP.
CANNOT INTERACT WITH
ALARM WATCH parameter
displayed
The monitor may not be configured
to allow interaction with remote
parameters.
Contact your system
administrator.
CANNOT deselect ALARM
WATCH
The Ultraview Central Monitor is
configured with Central Alarm
Watch Manager to prevent de-
selection of automatically alarm
watched beds.
Contact your system
administrator.