User's Manual

Ultraview Care Network
5-10
For the UCW and Ultraview 1700, the DISPLAY PRIORITIES key displays a list of
the currently active parameters, or all possible parameters, ordered by priority.
Touch LOCAL to display only this bed’s currently active parameters.
STORE — causes the currently displayed parameter priorities and colors to be
stored in memory.
RESTORE — causes the last stored local priority and color settings to be restored
from memory.
FACTORY DEFAULTS — causes the factory default priority and color settings to
be restored.
Data Shuttle Option
The Data Shuttle option enables you to transfer patient admit data (e.g., age,
gender, name, and BSA) and up to 24-hours of trend and episodic data from one
monitor to another using a 90496 Ultraview Command Module.
You can transfer data acquired through any Ultraview Care Network module or
Flexport interface. Before you shuttle data from one monitor to another, you must
first transfer the data from the source monitor into a 90496 module. When you
remove the module from the source monitor and insert it into the receiving
monitor, data can be transferred to the receiving monitor.
Transferring Data to the Module
If your monitor and module support the Data Shuttle option, the monitor displays a
TRANSFER DATA key in your ECG Setup menu. When you wish to transfer data
(for example when transporting a monitored patient from one unit to another),
touch the TRANSFER DATA key to initiate data transfer into the module. The
monitor will then display the message TRANSFERRING PATIENT DATA INTO
MODULE.
The time and date set on both the sending and receiving
monitors must be identical for the data transfer to be
successful.
The Data Shuttle option is only available in bedside or transport
monitors. Central monitors do not display keys and messages
related to data transfer.
If you insert a module that supports the Data Shuttle option into
a monitor that does not have the option, the monitor will not
display data transfer messages and keys.
To transfer data:
1 Touch ECG.
2 Touch SETUP.
3 Select TRANSFER DATA.
4 Remove module after DATA
TRANSFER COMPLETED is
displayed.