User's Manual

Ultraview Care Network
5-12
Retrieving Transfer Data
When you insert the module into the receiving monitor, the monitor displays the
RETRIEVE DATA and CANCEL TRANSFER keys and sounds a low priority alarm
tone every 5-seconds until you either cancel the data transfer or retrieve the data.
The monitor also displays the patient name associated with the data to be
transferred and, if applicable, the patient name associated with the data that
currently resides in the monitor.
When you touch the RETRIEVE DATA key, the monitor displays YES and NO
keys along with the following message:
YES purges the monitor’s data and retrieves data from the module.
NO cancels this action.
Touch NO to return to the previous screen. Touch YES to transfer the data from
the module into the receiving monitor and purge any data previously stored in that
monitor.
Figure 5-7: Retrieving Transfer Data display
Once you initiate data retrieval, the monitor removes all data transfer-related keys
from the ECG display zone and displays the message TRANSFERRING
PATIENT DATA INTO MONITOR. When the data retrieval into the monitor is
complete, the monitor displays the message DATA TRANSFER COMPLETED.
for one minute.
When you insert a module loaded with transfer data, you cannot purge a prior
patient's data (while data is being transferred) using the monitor’s Admit or
Discharge menus. If you attempt to purge patient data in this manner, the monitor
will display the following message:
Purge is not allowed during transport.
Canceling the Data Transfer After Module Insertion
To cancel a data transfer, touch the CANCEL TRANSFER key. The monitor
displays YES and NO keys along with the following message:
YES purges data previously transferred into the module.
NO cancels this action.
Touch YES to confirm the cancellation of the data transfer and reactivate the
TRANSFER DATA key in the ECG Setup menu. Touch NO to return to the
previous screen.
To receive transferred data:
1 Touch RETRIEVE DATA.
2 Touch YES.
CST
E
C
G
CANCEL TRANSFER
Module: Jones, B
Monitor: Adams, S
RETRIEVE DATA
To cancel data transfer:
1 Insert module into receiving
monitor.
2 Touch CANCEL TRANSFER.
3 Touch YES.