User's Manual

Temperature
16-5
Setting or Adjusting Temperature Alarm
Limits
High and low alarm limits can be set for T1 (T3), T2 (T4), and the temperature
difference (DT) between two temperatures.
The factory default settings for high and low temperature alarms are 37.5°C and
36.5°C. Refer to Default Alarm Limits on page 2-3 for details on system alarm
operation.
Recording Temperatures
You can send currently displayed temperature readings to a bedside recorder or
system printer. Refer to Printing on page 4-5 for an overview of system printing
information.
Restoring User-Defined Settings
With the Restore Settings feature, you can restore the user-configurable settings
to the defaults previously stored as user settings. User-configurable settings are
listed in Setting User-Defined Default Values on page 7-7.
!
RESTORE SETTINGS changes the user-configurable settings
for all parameters in the module.
To set or adjust alarm limits
1 Touch TEMP.
2 Touch ALARM LIMITS.
3 Select TEMP 1, TEMP 2,
TEMP 3, TEMP 4, or
DELTA TEMP.
4 Select ALARMS ON.
5 Select HI= or LO=.
6 Use arrow keys to adjust limit.
To print a strip chart of current
temperature readings:
1 Touch RECORD.
2 Touch TEMP while it flashes.
To restore user-defined settings:
Configurations with ECG
1 Touch ECG.
2 Touch SETUP.
3 Touch RESTORE SETTINGS.
4 Select YES.
To restore user-defined settings:
Configurations without ECG
1 Touch TEMP.
2 Touch RESTORE SETTINGS.
3 Select YES.