User's Manual

Digital Telemetry
17-17
SIGNAL INTERFERENCE
The signal interference message indicates, via the displayed triangle squelch
waveform, that an interfering signal has been detected. A signal can no longer be
detected because of interference from a stronger signal source lasting more than
0.5-seconds. A low priority alarm tone sounds whenever this message is
displayed in the waveform zone.
IS SIGNAL LOSS PERMANENT?
The permanent signal loss message indicates that no RF signal is being detected.
Multiparameter Telemetry (NIBP)
Setting Up NIBP Monitoring
Proper cuff selection and application is critical in ensuring the accuracy of NIBP
readings. To ensure proper cuff selection, first measure the circumference of the
limb at its midpoint. Match the limb measurement to the range of appropriate
circumferences (in centimeters) specified on each cuff. If the cuff bladder is too
wide for the patient, the reading will be falsely lowered; if it is too narrow, the
reading will be falsely elevated. Undersizing the cuff results in the greatest chance
of error, so a variety of cuff sizes should be available to accommodate your full
patient population.
Apply the cuff snugly. When the cuff is properly applied to an adult, you should be
able to insert one finger between the cuff and the arm. If you can insert two
fingers, the cuff is too loose, which may result in falsely elevated readings. Ensure
that the hose is not kinked when the cuff is applied.
During blood pressure measurement, the inflated cuff reduces blood flow to the
limb to which it is applied. Do not apply a cuff to a limb that has restricted blood
flow. Check the patient periodically.
!
Do not apply a blood pressure cuff to a limb being monitored
with a pulse oximetry sensor, because SpO
2
is affected during
NIBP readings. Avoid applying a cuff to a limb that has an
intravenous line in place. Do not apply a cuff to a limb that has
restricted blood flow.
Use only single hose cuffs to ensure proper operation.
Spacelabs Medical’s hoses are non-conductive with respect to
defibrillator discharge effects.