User's Manual

Ultraview Care Network
17-20
Figure 17-4: Common cuff hose positions
7. Connect the hose to the monitor.
8. To verify proper monitor operation, take one or more blood pressure readings.
Push the START/STOP key to begin a measurement.
9. The 90343 transmitter must be configured for use with the 90217 ABP monitor
by opening the battery compartment door, removing the battery, and setting
DIP switches 5 ON and 8 OFF. Refer to Figure 17-2 on page 17-12.
10. The 90478 receiver must be configured for operation with the 90343
transmitter and attached 90217 ABP monitor. Touch the ECG key to display
the main menu. Touch CHANNEL FORMAT, then touch NIBP ON. The
monitor will display the NIBP measurement in a numeric format in the display
zone. The values of the measurement are displayed as ??? until a valid NIBP
measurement has been taken.