User's Manual

17-24
NIBP Troubleshooting Guide
Clinical Situation Possible Cause Solution
No NIBP displays Adapter cable not inserted
correctly.
Remove and re-insert adapter cable.
NIBP not enabled on 90343 or
90478.
Enable NIBP function by setting transmitter
DIP switch 5 ON and setting DIP switch 8
OFF.
90217 ABP not properly initialized. Re initialize 90217 ABP using 90121.
90343 Low battery indicator
constantly illuminated.
Call qualified service person.
No NIBP readings can
be obtained
Incorrect or inoperative cuff in use. Replace with cuff known to be operative.
Tubing is kinked. Locate kink and straighten tubing.
Some arrhythmias (e.g., atrial
fibrillation and frequent ventricular
ectopy) may cause a single or
repeated failure to obtain a reading
(may be due to true beat-to-beat
variations in pressure).
Document arrhythmia if present, verify
pressure with another method, then follow
hospital procedure for care of this type of
patient.
Excessive patient motion or muscle
contractions associated with
shivering or severe pain.
Ensure that patient is quiet with minimal
movement during NIBP readings; minimize
patient’s shivering.
Blood pressure outside of
measurement range.
Verify extremely high or low pressure with
another method.
Intermittent or complete
failure to operate
90217 ABP error. Remove 90217 ABP from service; record
event code; and call qualified service person.
Apparent incorrect
value
Wrong size cuff for patient. Measure patient’s limbs at the midpoint;
match limb measurement to range specified
on cuff (undersizing the cuff results in the
greatest degree of error).
Cuff is damaged. Replace with good cuff.
Excessive patient motion, shivering
or severe pain.
Ensure patient is quiet with minimal
movement during NIBP readings; minimize
patient’s shivering.
False high readings may be the
result of venous congestion caused
by frequent readings.
Reduce frequency of readings.
Cuff too loose or positioned
incorrectly.
Tighten cuff or reposition appropriately.