User's Manual

18-11
Invasive Pressure Troubleshooting Guide
Clinical Situation Possible Cause Solution
Intermittent or no operation Module error. Call a qualified service person.
No pressure key appears Module not inserted correctly. Reinsert the module.
Transducer not connected. Reconnect the transducer.
Numeric display does not
settle
Respiration artifact too high. Select ART REJ ON.
Pressure display disappears Cable disconnected from the
module.
Reconnect cable.
Cable disconnected from the
transducer.
Reconnect cable.
Pressure shows NOT
ZEROED
Pressure has not been zeroed. Zero pressure with the ZERO key
after opening transducer to air.
Shows constant pressure Stopcock(s) are positioned
incorrectly.
Reposition stopcock(s) to connect
the patient to the transducer
(a waveform will appear on the
screen).
ZERO REJECTED message
appears
Stopcock(s) are positioned
incorrectly.
Reposition stopcock(s) to open the
transducer to air. Zero pressure with
the zero key.
Still unable to zero. Follow transducer manufacturer’s
instructions to correct the problem.
CHECK CATHETER key
appears on the monitor
Stopcock(s) are positioned
incorrectly.
Reposition stopcock(s) to connect
the patient to the transducer
(a waveform will appear on the
screen).
UA catheter has become dislodged
or occluded.
Check UA catheter.
Upon completion of the above, touch
the CHECK CATHETER key to
silence the alarm.