User's Manual

9-16
Arrhythmia Troubleshooting Guide
Clinical Situation Possible Cause Solution
Abnormal beat not
detected
Module receiving inadequate signal;
NOISY SIGNAL or ECG VOLTAGE
TOO LOW messages appear. Noise
level is over allowable range, or
signal level is below QRS detection
threshold.
Make the necessary adjustments to
restore good signal.
Check all leads to determine a better
monitoring lead or select another
lead.
Remove the cause of the noise.
Some beats not recognized as
morphologically different from the
learned dominant beat.
No action required; some events that
may be diagnosed as abnormal by a
skilled clinician may not meet the
module’s criteria for abnormality.
System has not seen 5 abnormals or
3 couplets of like morphology to
generate a class.
No action is required.
Arrhythmia detection is not enabled. Enable arrhythmia detection through
the ECG Setup menu.
False Alarms NOTE: Careful attention to good monitoring technique, especially setup, will keep false
alarms at a low level. If false alarms do occur, however, check the following.
Noise on the signal caused by poor
electrode application is the most
common cause of false alarms.
Remove the cause of the noise.
Deactivate alarm for the classes that
fill up with repetitive artifact; do not
clear these classes.
Multiple abnormal classes with atrial
fibrillation or flutter waveforms.
Merge the class with the dominant or
another abnormal class.
Deactivate alarm for these classes,
or consider deactivating the ABN
PER MIN alarm.
Atrial fibrillation continually triggering
TACH alarm.
Consider deactivating TACH alarm.
Limits set too close to patient’s heart
rate.
Check and adjust the alarm limits.
Amplitude of ECG signal has
dropped below threshold of R-wave
detector.
Reposition electrodes and relearn
patient’s rhythm.