ST Analysis Troubleshooting Guide
Clinical Situation Possible Cause Solution
No access to ST analysis
functions
The system must learn the ST
segment level before it can provide
access to ST analysis functions.
Wait until the system analyzes
sufficient QRS complexes to
calculate the ST segment level
(approximately one minute).
ST = ?? is displayed Infrequent occurrence of dominant
beats.
ST analysis not performed on paced,
premature dominant or abnormal
beats.
ECG amplitude may be insufficient
to detect QRS complexes.
Check QRS amplitude.