User's Manual
Ultraview Care Network
11-4
Display Detail
The 12-lead report display shows 2.5-seconds of waveform data per lead.
Presentation of the leads may be in a standard format (refer to Figure 11-1) or the
Cabrera format. When analysis is complete, measurements and diagnostic
statements are displayed above the waveform data (requires option D).
Figure 11-1: 12-lead report display
Measurement and interpretation data (requires Option D)
ECG traces for 12 leads (2.5-seconds/lead)
Initiating a 12-Lead Report
You can acquire a 12-lead report by touching STAT REPORT. Initiating a STAT
REPORT will not affect the automatic schedule.
!
• When monitoring paced patients, ECG processing and the
pacepulse detection function will be suspended for 10-seconds
whenever acquisition of an ECG report is in process.
Temporarily suspending these functions allows the actual
pacepulse to be displayed/printed in the 12-lead report without
interfering with arrhythmia analysis.
• If the STAT REPORT key is labeled LEADS OFF, one or more
of the 12 leads is disconnected. Check all electrode
connections before proceeding.
Vent. rate: 60 BPM SINUS BRADYCARDIA
PR interval: 162 ms NORMAL ECG
QRS duration: 88 ms
QT/QTc: 360/360 ms
P-QRS-T axes: 50 44 51
I AVR V1
V4
II AVL V2
V5
III AVF V3
V6
To initiate a 12-lead ECG report:
1 Touch ECG.
2 Touch REVIEW.
3 Touch REPORT REVIEW.
4 Touch STAT REPORT.