Display Detail
The 12-lead report display shows 2.5-seconds of waveform data per lead.
Presentation of the leads may be in a standard format (refer to Figure 11-1) or the
Cabrera format. When analysis is complete, measurements and diagnostic
statements are displayed above the waveform data (requires option D).
Figure 11-1: 12-lead report display
Measurement and interpretation data (requires Option D)
ECG traces for 12 leads (2.5-seconds/lead)
Initiating a 12-Lead Report
You can acquire a 12-lead report by touching STAT REPORT. Initiating a STAT
REPORT will not affect the automatic schedule.
!
When monitoring paced patients, ECG processing and the
pacepulse detection function will be suspended for 10-seconds
whenever acquisition of an ECG report is in process.
Temporarily suspending these functions allows the actual
pacepulse to be displayed/printed in the 12-lead report without
interfering with arrhythmia analysis.
If the STAT REPORT key is labeled LEADS OFF, one or more
of the 12 leads is disconnected. Check all electrode
connections before proceeding.
Vent. rate: 60 BPM SINUS BRADYCARDIA
PR interval: 162 ms NORMAL ECG
QRS duration: 88 ms
QT/QTc: 360/360 ms
P-QRS-T axes: 50 44 51
I AVR V1
V4
II AVL V2
V5
III AVF V3
V6
To initiate a 12-lead ECG report:
1 Touch ECG.
2 Touch REVIEW.
3 Touch REPORT REVIEW.
4 Touch STAT REPORT.