Turning Respiratory Waveform On/Off
You may choose to turn the respiration waveform OFF and display only the
respiration numeric values.
The factory default setting for respiration waveform is ON.
Adjusting Waveform Size
You can increase or decrease the size of the respiration waveform display. This
changes the displayed size only, and does not affect the signal gain or breath
detection sensitivity.
Adjusting Sweep Speed
The sweep speed determines the rate at which the respiration waveform moves
across the screen. Sweep speed selections are 25, 12.5 (factory default), 6.25, or
1.56 mm/sec.
Adjusting Respiration Tone
You can select an audible tone to sound with each respiratory cycle. The tone
volume is adjustable or can be disabled.
The factory default setting for tone is OFF.
Selecting Respiration Leads
Respiration lead selections are RL-LA, RA-LA, RA-LL, and RL-LL.
RA-LA represents the line of maximum respiratory effort in adults and chest-
breathing infants.
RA-LL represents the line of maximum respiratory effort in abdominal-
breathing infants.
The factory default setting for respiration lead selection is RA-LA.
!
When a good respiration signal cannot be attained from the
selected lead, the message LOSS OF SIGNAL is displayed and
??? appears in place of the respiratory rate.
To turn the respiration waveform
display ON or OFF:
1 Touch RESP.
2 Select WAVEFORM ON/OFF.
To adjust respiratory waveform
size:
1 Touch RESP.
2 Touch SIZE.
3 Adjust waveform size using
SIZE or SIZE ↓.
To select a respiration sweep
speed:
1 Touch RESP.
2 Touch SWEEP SPEED.
3 Touch the appropriate mm/sec
key.
To adjust the respiration tone
volume:
1 Touch RESP.
2 Touch RESP TONE.
3 Select TONE ON.
4 Adjust tone volume using
VOLUME or VOLUME↓.
To select respiration leads:
1 Touch RESP.
2 Touch LEAD SELECT.
3 Select the appropriate lead
configuration.