Error Messages
When error messages are displayed, the saturation value and pulse rate
immediately change to ??? and an alarm may occur if your module is configured
to do so. If it occurs, this alarm does not begin until after the message alarm delay
time has elapsed.
If your module has been configured for an alarm using the Module Configuration
Manager feature, the parameter key will flash red for high priority alarms or yellow
for medium and low priority alarms. Refer to Module Configuration Manager on
page 7-5 for more information.
Telemetry products use different text for their SpO
2
error messages than non-
telemetry products. Table 1 serves as a key for interpreting the telemetry
messages. Telemetry products also display their error messages within ECG’s
display zone, so the following ECG alarm messages take priority over other SpO
2
messages.
LEADS OFF
NOISY SIGNAL
ECG ALARMS SUSPENDED
WARNING:
Error messages indicate a problem or condition which may
affect accurate monitoring values. Do not ignore these
messages. Correct any fault before continuing.
Table 1: Telemetry Message Equivalent
SpO
2
Message
Equivalent Telemetry SpO
2
Message
ADAPTER DISCONNECTED – CHECK
CONNECTION AT MODULE
SPO2 UNAVAILABLE
FAULTY SENSOR – REPLACE SENSOR SPO2 FAULTY SENSOR
SENSOR DISCONNECTED – CHECK
CONNECTION AT ADAPTER CABLE
SPO2 SENSOR
DISCONNECTED
SENSOR OFF PATIENT – CHECK
CONNECTION AT PATIENT
SPO2 SENSOR OFF
PATIENT
INSUFFICIENT SIGNAL – REPOSITION OR
REPLACE SENSOR
SPO2 INSUFFICIENT
SIGNAL
AMBIENT LIGHT INTERFERENCE – COVER
SENSOR AREA
SPO2 AMBIENT LIGHT INTF.
NOISY SIGNAL SPO2 NOISY SIGNAL
LOW SIGNAL STRENGTH – REPOSITION
OR REPLACE SENSOR
SPO2 INSUFFICIENT
SIGNAL