NOISY SIGNAL
Sensor signal disturbed by motion or other interference. Eliminate sensor
movement. Message disappears when the value is obtained.
Sensor placement adjacent to power cords or other electrically noisy devices.
Move the noisy device or move the sensor to another site.
If the message persists, replace the sensor.
LOW SIGNAL STRENGTH — REPOSITION OR REPLACE SENSOR
When this message appears the saturation and pulse rate continue to be
displayed; however, the Sensorwatch flashes as a indication of a possible error
condition.
Insufficient blood flow between the sensor light emitter and detector. Move
the sensor to an area of higher perfusion.
Poor sensor application. Reposition to place active components closer to skin
or locate to a better perfused site.
Sensor site is below the blood pressure cuff. Move to another site.
If the message persists, replace the sensor.
Sensors
For further information about biocompatibility or sensor disposal, refer to the
Spacelabs Medical Product Specification sheets and to the manufacturer's