User's Manual

14-13
SpO
2
Troubleshooting Guide
Clinical Situation Possible Cause Solution
No SpO
2
parameter key
is displayed
Module not inserted correctly. Remove and reinsert the module.
Adapter cable not connected to
module properly.
Correctly connect the adapter cable.
Sensor not connected to adapter
cable.
Correctly connect the sensor.
SpO
2
is not enabled at the 90343
transmitter.
Call a qualified service person to
check transmitter DIP switch 1 and 2.
SpO
2
is not enabled at the 90478
receiver.
Call a qualified service person to set
transmitter DIP switch 8 to OFF.
SpO
2
value displays ??? Sensor not connected to patient. Reattach the sensor.
Excessive patient motion. Urge patient to remain still while
reading is in progress.
Module is in the initialization phase
(the first 15-seconds after sensor
application).
Wait until the initialization is
complete.
Adapter cable not connected to the
module properly.
Correctly connect the adapter cable.
Sensor not connected to the adapter
cable.
Correctly connect the sensor.
Low battery indicator constantly
illuminated.
Call a qualified service person.
Low signal strength Sensor placement not optimum. Move the sensor to a site which has
better perfusion.
Align the LED with the sensor photo
detector.
Sensor is placed below blood pressure
cuff.
Move the sensor to an alternate limb.
Intermittent or complete
failure to operate
Module error. Call a qualified service person.