User's Manual

Fetal Monitoring
15-15
Type 2
An alert will activate if:
•Group H is less than Group F; and
•Group G is less than Group E; and
•Group F is less than Group D; and
•Group E is less than Group C; and
•Group H is below Group C by more than 30 bpm.
Figure 15-4: Type 2 bradycardia alert
Type 3
An alert will activate if the average of Groups G and H is lower than the average of
Groups A, B, C, and D by more than 40 bpm.
Figure 15-5: Type 3 bradycardia alert