User's Manual

Fetal Monitoring
15-23
Module Configuration Manager
The Module Configuration Manager feature provides the capability to define and
store all user-configurable options for the fetal parameter. Once configured, these
settings control its operation whenever the monitor is powered ON.
Display Detail
The User Settings screens list all user-configurable options. From these screens
you can change and store the options in your monitor.
Figure 15-15 illustrates the User Settings screen which is used to change and
store fetal rate alarm limits.
The parameter configuration table in this chapter lists all available options, the
available user settings, and the factory defaults for each parameter.
Figure 15-15: User settings
CURSOR and CURSOR keys scroll through the two columns of
parameters on the screen.
and keys scroll the list of valid settings for the highlighted parameter.
PREV PAGE key returns to the previous screen.
NEXT PAGE key advances to the next screen.
PRINT PAGE key prints the current screen.
STORE key stores the currently selected settings as defaults.
PARAMETER DESCRIPTION VALUE PARAMETER DESCRIPTION VALUE
FETAL RATE ALERT OFF ECG RATE ALARMS (INFANT) NO SHIFT
HIGH RATE 180 TOCO SENSITIVITY MEDIUM
LOW RATE 84 ECG PLOT ON
FETAL ECG WAVEFORM ON ECG EDIT OFF
FECG WAVEFORM SIZE 1.00 EXTERNAL DEVICE NONE
FECG SWEEP SPEED 25 PRINT PERIOD 10
Enable/Disable Fetal Rate Alert

STORE
SELECT PARAMETER
CURSOR
CURSOR
CHANGE VALUE
PAGE X OF X
PRINT
PAGE
NEXT
PAGE
PREV