User's Manual

Ultraview Care Network
19-10
Editing Vital Sign Values
Select a row of data in the hemodynamic calculations table to turn the vital signs
keys ON.
Default values are supplied if you touch the VITAL SIGNS key and no prior value
is available. When you edit a value, the system recalculates the hemodynamic
calculation values using the new vital sign value.
A CVP value is used in calculations if it is available. If only a RAP value is
available, then the RAP value is used in place of CVP. If neither CVP nor RAP
pressure is available, the system cannot automatically calculate SVR, SVRI,
RVSW, or RVSWI. A CVP or RAP value can be manually entered using this edit
process.
A PCWP value is used in calculations if the last stored PCWP value is less than
15-minutes old. If no such PCWP value is present, an LAP value is substituted. If
neither PCWP nor LAP values are available, the system cannot automatically
calculate PVR, PVRI, LVSW, nor LVSWI and the monitor displays the message
NO PCWP VALUE AVAILABLE WITHIN THE LAST 15-MINUTES. You can add a
PCWP value manually using this edit process.
View Additional Table Data
Five sets of values display at any one time. Additional sets of values can be
displayed by scrolling through the data.
Thirty sets of values are saved in the monitor’s hemodynamic calculations table
so you can remove and re-insert the module without losing hemodynamic values.
You can erase these values by discharging a patient or by powering the monitor
OFF.
Table 2: Hemodynamic and Vital Signs Values
Label Name Units
Default
Value
Valid
Range
HR Heart Rate beats/min 70 0 - 300
MAP Mean Arterial Pressure
mmHg
kPa
80
10.7
0 - 300
0.0 - 40.0
CVP Central Venous Pressure
mmHg
kPa
10
1.3
0 - 99
0.0 - 13.2
MPA
Mean Pulmonary Artery
Pressure
mmHg
kPa
15
2.0
0 - 99
0.0 - 13.2
PCWP
Pulmonary Capillary
Wedge Pressure
mmHg
kPa
10
1.3
0 - 99
0.0 - 13.2
!
To ensure that a RAP/CVP value can be registered,
immediately reopen the stopcock to the patient after you inject
the bolus so that flow is reinstated.
To edit vital sign values:
1 Touch CO.
2 Touch CARDIAC OUTPUT.
3 Touch CALCS.
4 Touch DAY/TIME in the row
you wish to select.
5 Touch VITAL SIGNS.
6 Select the vital sign you wish to
edit.
7 Use arrow keys to edit the
displayed value.
8 Press ENTER.
To view additional sets of data:
1 Touch CO.
2 Touch CALCS.
3 Touch SCROLL UP to scroll the
data up one row, touch
SCROLL DOWN to scroll the
data down one row.