User's Manual

Calculations
25-11
Hemodynamic Calculations require several specific pressure values, including
CVP, MAP, and PCWP. CVP and MAP are monitored continuously; PCWP is an
episodic value. A data substitution may occur if any of these pressure values are
not available when you touch NEW ENTRY. When a substitution occurs, one of
the error messages listed here displays on the message line.
Warning...RAP has been substituted for CVP.
A continuous RAP value is used instead of the continuous CVP value if CVP
is not available.
Warning...NIBP has been substituted for MAP.
An episodic mean NIBP value which is less than 15-minutes old is used
instead of MAP if the continuous MAP value is not available.
Warning...LAP has been substituted for PCWP.
A continuous LAP value is used instead of the episodic PCWP value if the
PCWP value is not available or is more than 15-minutes old.
Table 3: Hemodynamic Output
Label Name Units
CI Cardiac Index L/min
2
SV Stroke Volume ml/beat
SVI Stroke Volume Index ml/beat/m
2
SVR Systemic Vascular Resistance dynes x sec/cm
5
SVRI Systemic Vascular Resistance Index (dynes x sec/cm
5
) x m
2
PVR Pulmonary Vascular Resistance dynes x sec/cm
5
PVRI Pulmonary Vascular Resistance Index (dynes x sec/cm
5
) x m
2
LVSW Left Ventricular Stroke Work g x min/beat
RVSW Right Ventricular Stroke Work g x min/beat
LVSWI Left Ventricular Stroke Work Index g x min/beat/m
2
RVSWI Right Ventricular Stroke Work Index g x min/beat/m
2
Table 4: Hemodynamic Equations
BSA = Ht
0.725
x Wt
0.425
x 0.007184
CI = CO/BSA
SV = (CO/HR) x 1000
SVI = SV/BSA
SVR = 79.9 x [(MAP-CVP)/C
O]
SVRI
= 79.9 x [(MAP-CVP)/CI]
= 79.9 x [(MAP-CVP)]/[CO/BSA]
= 79.9 x [(MAP-CVP)]/[CO x 1/BSA]
= 79.9 x [(MAP-CVP)/CO] x BSA
= SVR x BSA