Adjusting the Table Range
Use the SCROLL and PAGE keys in the Titration Table menu to select the range
of data displayed in the titration table. The default range is 1-100.
Printing
Touch the PRINT key to print the currently displayed page of the Calculations
table. The complete table prints on a 4-channel recorder. On a 2-channel
recorder, the top half of the table prints first followed by the bottom half
Recalculations
You can recalculate all Drug Dosage inputs except for the amount/hour entries
shown for DOSE and DOSE/WT. In most cases, changing one parameter
automatically recalculates and redisplays other parameters under that DRUG key.
Table 16 shows adult patient recalculation rules, except where noted in AMOUNT
and VOLUME. When TITRATE on the MIX/TITRATE key is selected, DOSE,
DOSE/WEIGHT, and FLOW RATE results are the same for Adult and Neonatal.
CAUTION:
If any value in the Drug Calculation table is over its
designated range, ++++.++ will be displayed for values of
AMOUNT, CONC or DOSE, or RATE and the Titration Table
becomes inaccessible.
To print calcs data:
1 Access Local or Remote Drug
Calcs.
2 Touch PRINT.