User's Manual

Ultraview Care Network
29-4
The sample rate refers to the frequency of data sampling and can be set to time
intervals ranging from 5-seconds to 60-minutes. The new sample rate takes effect
at once.
The data logger output requires a serial cable that swaps pins (refer to Table 1).
The serial cable must be properly connected at both ends and the data logger
output enabled.
!
The option is only available on bedside monitors.
Ensure that any cables or other computer or communications
equipment connected to the bedside monitor comply with
applicable medical standards.
The combination of PDL baud rate and sample rate must be
selected carefully. Slow baud rates, in combination with
frequent data transmissions, may cause loss of data. This is
especially true if a large amount of patient vital sign data is
being monitored. However, some devices cannot accept data at
high baud rates. This may limit the range of sample rates at
which your device can accept data readings from the monitors.
Table 1: Serial Cable Pinout
Input Output
23
32
45
54
To set up Patient Data Logger
(the user must have System
Administrator access):
1 Touch MONITOR SETUP.
2 Touch PRIVILEGED ACCESS.
3 Enter biomed password.
4 Touch SERIAL PORTS and
select a port.
5 Touch ASSIGNMENT.
6 Touch DATALOGGER.
7 Touch PREVIOUS MENU.
To set serial settings:
8 Touch SETTINGS.
9 Touch the desired setting
key(s) to display desired
settings.
10 Touch NORMAL SCREEN to
effect changes.