Input leakage current for all patient input channels is less than 10 µA making ECG
units suitable for direct cardiac application. The maximum non-destructive voltage
that can be applied to any input or output connector on the monitors, modules or
printers is +5 V. All signal inputs or outputs are for exclusive connection to
equipment specified by Spacelabs Medical.
Disposal of these devices and all accessories must be in accordance with local
and federal laws.
When interconnecting equipment, avoid using optional battery operation. If such
operation cannot be avoided, the battery-operated instrument must be grounded
using the ground terminal on the rear panel or the protective ground (earth) in the
power cord.
Power Cord:
3-wire, 18-gauge, hospital grade, or
3-wire, 0.75mm
2
, European harmonized.
Plug:
3-terminal polarized, with protective ground.
Table 1: Ultraview Care Network Products Power Rating
Product Frequency Electrical Rating Fuse Rating
90367/90369 50/60 Hz
100-240VAC
2.0-1.0A
N/A
90385/90363 50/60 Hz
2.5A/100-120V
1.3A/220-240V
2-T3A/250V
2-T3.15A/250V
90386 50/60 Hz
1.35A/100-120V
0.8A/220-240V
2T2A/250V
2-T1.25A/250V
90479-A 50/60 Hz
2A/115V
1A/230-240V
2-T2.5A/250V (Slow)
2-T1.6A/250V
90364/90387/90491/
90499/94000
50/60 Hz
100-240V
2.0-1.0A
N/A
WARNING:
To protect against electrical shock, proper grounding is
essential.
If a monitor is in battery mode during defibrillation, the
monitor must be properly grounded using the ground
terminal provided on the rear panel or by using the
protective ground in the power cord.
If the integrity of the external protective earth conductor
is in doubt, the equipment must be operated from its
internal power source (if applicable).
WARNING:
Do not use a 3-to-2 plug adapter.
Ground terminal of the plug is connected directly to the
frame of the instrument. Any interruption of the grounding
connector can create an electric shock hazard.