User's Manual

Ultraview Care Network
31-4
Equipment Maintenance Requirements
Corrective or maintenance procedures must be performed by qualified personnel.
Periodic maintenance procedures are required every 12 months to verify that:
the equipment is physically sound.
resistance between the chassis ground connector on the rear panel and the
protective ground of the mains input is not greater than 0.1 ohm.
isolation resistance between ground and mains is greater than 2 meg-ohm.
While the equipment is operating normally verify that:
chassis leakage current is less than 100 µA.
patient leakage current is less than 10 µA (Type CF),
100 µA (Types B and BF).
While the equipment is operating in single fault condition verify that:
chassis leakage current is less than 300 µA (100-120 V), 500 µA (220-240 V);
patient leakage current is less than 50 µA (Type CF),
500 µA (Types B and BF).
Under non-optimal environmental conditions or periods of intense use, more
frequent checks are recommended.
If the equipment has been dropped, abused or damaged in any way (if the monitor
or module becomes wet, for example), a qualified service person must verify that
the unit is working correctly and that all safety features are intact.
Visually inspect all patient cables each time the unit is used. Check for worn or
damaged plastic covering, frayed or broken wires, cracked connections or any
other signs of damage. Do not use cables which exhibit obvious damage.
Ultraview Care Network Module
Compatibility
If any function (NIBP, ECG, SpO
2
, etc.) on your system does not contain a feature
described in this manual:
Your product may contain an earlier version of software. Contact your
customer service representative and refer to the original documentation that
accompanied your system.
Your system configuration may be different from that described in this
manual. Refer to notes in this manual describing features where system
configuration is likely to impact the available features.
Before moving a Ultraview Care Network module from one network to another, be
certain that the module software version is compatible with that required by the
second network. If in doubt, have a qualified service person verify compatibility
between the module and the network.