User's Manual

Cleaning and Sterilization
32-5
90518 Multigas Analyzer
Exterior
The exterior of the 90518 should be cleaned and disinfected as necessary. Use a
clean cloth or swab dampened in a solution of warm water and mild detergent.
Squeeze the cloth out thoroughly before use. Disinfect the unit with Hemosol or
similar disinfectant.
Disposable and reusable patient accessories are available for the 90518.
Disposable accessories are for single-patient use only and must not be
sterilized or cleaned for reuse on other patients. Refer to the instructions
provided with each patient accessory to determine if the accessory may be
cleaned and reused on the same patient.
Reusable accessories can be used on multiple patients after cleaning and/or
sterilizing. Refer to the instructions provided with these accessories for
details.
Instructions provided with patient accessories may contain warnings regarding
their use. Read these instructions carefully prior to use.
Fan Grill
The analyzer has a fan grill located on the rear panel. It must be kept free from
dust and other contaminants.
WARNING:
Always turn the unit off and unplug the power cord before
cleaning to protect against electrical shock.
CAUTION:
Do not allow liquid to enter the interior of the 90518. If this
should occur, check the unit for proper operation and
verify its performance accuracy prior to reuse.
!
The Spacelabs Medical sampling lines are for single-patient use
only. Cleaning deteriorates the properties of the sampling line,
resulting in slower response time and more frequent occlusions.