33-2
REMOTE BED SELECT FEATURE IS IN USE BY
ANOTHER APPLICATION
The Remote Bed Select window is in use by another application.
Only one application may use this window at one time. Touch the
application which is using the Remote Bed Select feature and touch
PREVOUS MENU or NORMAL SCREEN to abort the selection.
The window will now be available for use.
UNABLE TO RECORD THE REQUESTED ALARM
CHANNEL(S)
Unable to perform the requested operation. Verify that a recorder is
selected and operational (has paper) and try again.
UNABLE TO RECORD THE REQUESTED
CHANNEL(S)
Unable to perform the requested operation. Verify that a recorder is
selected and operational (has paper) and try again.
PARAMETERS NOT AVAILABLE FOR
RECORDING
Unable to perform the requested operation. Verify that a recorder is
selected and operational (has paper) and try again.
CAUTION - MONITOR SETTINGS MAY HAVE
CHANGED
The battery for non-volatile memory has failed. That memory resets
the monitor to the default settings and the default settings may not
match hospital preferences. Contact your hospital biomed to check
Biomed and Clinical Menu settings and Alarm Watch setup.
PAPER OUT There is no paper in the fetal chart recorder. Load paper.
RECORDER OFF
Indication that the fetal chart recorder is powered OFF. Power ON
fetal chart recorder.
“CHECK CABLE”
“NO CONNECTION”
Invalid CIM address. Verify that the CIM cable is properly
connected to the monitor and the wall outlet.
“CHECK CABLE” then “OLD XXX NEWXXX”
then “ID CHANGE!” ALTERNATELY
The CIM ID has changed unexpectedly during operation. Verify that
the CIM cable is properly connected to the monitor and the wall
outlet.
ADMIT PATIENT
No patient name has been entered. Enter patient name into
BirthNet or standalone monitor.
Message Cause/Action