User's Manual

21-16
Problem or
Message
Probable Cause(s) Suggested Response
Alarm
Sounds
Low or high CO2 values
are observed
An airway adapter and/or sensor
calibration is needed.
Perform a zero and a
reference calibration, make
sure an appropriate airway
adapter is selected, and
perform an airway adapter
calibration. If problem persists,
return to factory for repairs.
No
ADAPTER CALIBRATION
ABORTED
The airway adapter calibration
was aborted by the user.
Wait for the message to go
away.
No
O2 ZERO COMPLETE The zero calibration of the O
2
sensor was completed
successfully.
Wait for the message to go
away.
No
O2 ZERO REJECTED
Zero value too high
O
2
cell was connected during
zero.
Disconnect O
2
cell and repeat
the O
2
zero.
No
21% O2 SPAN
COMPLETE
The span calibration of the O
2
sensor in room air was completed
successfully.
Wait for the message to go
away.
No
O2 SPAN REJECTED -
Span value out of range
O
2
cell was not functioning during
the room O
2
span.
Replace O
2
cell perform room
O
2
span. If problem persists,
return to factory for repairs.
No
O
2
cell was not in room air. Place the O
2
sensor in room
air and away from any source
of higher level oxygen and
perform room O
2
span. If
problem persists, return to
factory for repairs.
No
100% O2 SPAN
COMPLETE
The span calibration of the O
2
sensor using 100% oxygen was
completed successfully.
Wait for the message to go
away.
No
100% O2 SPAN FAILED -
O2 cell bad or 100% O2
not present
O
2
cell is wearing out. Replace O
2
cell and perform
100% O
2
span. If problem
persists, return to factory for
repairs.
No
100% oxygen was not present in
the O
2
sensor airway adapter.
Make sure 100% O
2
is present
in the O
2
sensor airway
adapter and perform 100% O
2
span. If problem persists,
return to factory for repairs.
No