Clinical Situation Possible Cause Solution
Alarm
Sounds
HIGH VALUES
OBSERVED
A span calibration was
performed while there
was a leak in the
sample circuit.
Check connections and check for a
broken water trap. Perform leak test.
Verify the anesthesia delivery
devices and vaporizer calibrations.
Refer to the 90518 Multigas
Analyzer Service Manual
(P/N 070-0643-xx) for details.
Assure vaporizer accuracy.
No
LOW VALUES OBSERVED A leak exists at the
sample line
connection, a water
trap is broken at the
luer connection, or a
gas analyzer filter is
leaking.
Check connections and check for a
broken water trap. Perform leak test.
Verify the anesthesia delivery
devices and vaporizer calibrations.
Refer to the 90518 Multigas
Analyzer Service Manual
(P/N 070-0643-xx) for details.
No
Assure vaporizer accuracy. No
Minimum waveform
deflection and very low or
zero numeric values when
breathing normally
through a new or unused
sample line
Clogged or missing gas
analyzer filter.
Replace or install gas analyzer filter
underneath the door above the
water trap.
No
A large leak in the
sample circuit.
Check connections and perform a
leak test. Refer to the 90518
Multigas Analyzer Service Manual
(P/N 070-0643-xx) for details.
No
MIX
NOTE: This message
appears in agent
numerics area.
Agent mixture
detected.
The analyzer is not designed to
simultaneously measure multiple
anesthetic agents. Inaccurate
numeric concentration values may
be displayed when two or more
agents are mixed.
No
The displayed value of the mixed
condition is based on the combined
value of agents using the coefficient
factors of the last detected or
selected agent type.
No