User's Manual

Ultraview Care Network
23-4
Figure 23-1: Module, DSC, PIC, and sensor
CAUTION:
Place the sensors as far away from the electrosurgery site
as possible because considerable high frequency current
may flow into the electrodes. This can cause patient burns,
especially if a defect is present in the neutral cable of the
electrosurgical unit. Placing electrodes too close to the
electrosurgery site can also cause a noisy EEG trace.
Do not allow conductive parts of sensors or connectors,
including the neutral electrode, to contact other conductive
parts, including the ground.
Detach all connectors and cables by grasping the
connectors and pulling them straight out. Do not detach
connectors and cables by pulling on the cables
themselves.
BIS module
digital signal converter (DSC)
patient interface cable (PIC)
BIS sensor