User's Manual

BIS
23-5
Digital Signal Converter
The digital signal converter (DSC) receives the EEG input from the patient and
converts it to digital information that is processed by the module.
Figure 23-2: Digital signal converter
The DSC plugs into the front of the module as shown in Figure 23-1.
Patient Interface Cable
The patient interface cable (PIC) is attached to the three-foot cable on the DSC
and connects to the BIS sensor as shown in Figure 23-1.
To attach the PIC to the BIS sensor:
1. Remove the BIS sensor from its packaging.
2. Locate the small, raised arrows on both the light-blue end of the PIC connector
and the BIS sensor connector.
Plug the BIS sensor into the PIC so that the arrows are on the same side and
point toward each other. You may attach the BIS sensor to the PIC cable either
before or after the sensor is placed on the patient.
!
•The PIC is not a disposable cable. Do not dispose of it when
you dispose of the attached BIS Sensor.
A cable tie is provided with the PIC to secure it to the DSC
cable to prevent accidental disposal of the PIC.
top view side view bottom view
end view