User's Manual

BIS
23-13
BIS Main Menu
The Main menu is the initial menu that displays when you touch the BIS key.
Refer to the following sections for descriptions of each key:
ALARM LIMITS Key on page 23-13
SETUP Key on page 23-14
TEST ELECT Key on page 23-15
PRINT Key on page 23-16
START/STOP PROCED Key on page 23-16
ALARM LIMITS Key
Touch ALARM LIMITS on the Main menu to display the Alarm Limits menu.
This menu turns the BIS alarms ON or OFF, sets the upper and lower BIS alarm
limits, and turns the Caution alarm ON or OFF.
Setting Alarm Limits
To raise or lower each alarm limit setting, first touch the HI or LO key, then touch
the up or down arrow key. Each touch of an arrow key raises or lowers the alarm
limit by five units. There are separate values for the low limit and the high limit.
Caution Alarm
Touch CAUTION ON to enable the caution alarm.
When the caution alarm is enabled, you will be warned when the BIS value is
within five units of either alarm limit.
When the BIS value is within the caution zone, the BIS number flashes.
!
In the remote view, only the ALARM LIMITS, SETUP, and
PRINT keys are active. All other keys are disabled.
!
The alarm limits appear on the HI and LO keys as they are
changed.
!
No tone is emitted for the caution alarm.