User's Manual

Introduction
071-0774-00 Rev. A 1-9
Cleaning
Clean the transmitter after each use. The transmitter does not require any
preventive maintenance other than cleaning.
To clean the transmitter, use only the following agents.
Mild soap and water solution
U.S. Pharmacopoeia (USP) green soap
Sodium hypochlorite solution (1:10 dilution of household bleach in water)
Phenolic germicidal detergent solutions (1% aqueous solutions)
Isopropyl alcohol solution (70%)
Assigning a Telemetry Channel
Telemetry transmitters have preassigned channel frequencies. This channel
number is identified on the back of the case and cannot be changed. To receive
this telemetry channel, one of the receivers in the telemetry receiver housing must
be tuned to its assigned frequency.
Tuning a Receiver for a Bedside
The central monitor must be tuned by a qualified service person, but the bedside
monitor may be tuned using the ECG TM SETUP menu. You may use this menu
to tune the receiver module to the pre-assigned channel frequencies on the
telemetry transmitter.
!
Tuning telemetry receiver modules to transmitter channels at
the central monitor must be done by a qualified service person.
Your central monitor can be configured to remember beds that
are assigned to individual telemetry channels using the Module
Configuration Manager feature. These beds are permanently
assigned until you unassign or reassign them. Refer to the
Module Configuration Manager chapter of the UCN Operations
Manual.
!
The module default is set for North America using UHF band
operation. For alternate band operation, or if operating in
another country, you must select the appropriate frequency
band using the Module Configuration Manager.
To set up the central for ECG (if
bed name not remembered):
1 Touch key label that matches
transmitter's frequency
2 Select bed/room number for
transmitter channel
To set up the central for ECG
(MPT=OFF) (UCW and 1700):
1 Touch MONITOR SETUP
2 Touch SCREEN FORMAT
3 Select subnet and bed/room
number
4 Select ECG and then desired
zone
To tune a receiver module at
bedside:
1 Touch ECG
2 Touch SETUP
3 Touch TM SETUP
4 Touch SET TM CHANNEL
5 Select the digit to change. Use
the keys to select the value
for that digit
6 Repeat for all digits as
necessary
7 Touch STORE