User's Manual

Ultraview Digital Telemetry
2-8 071-0774-00 Rev. A
Two Operational Modes
Two operational modes are available with the 91341 dual-lead transmitter:
the standard mode that offers a choice of one V lead (V1-6), plus lead II; or
the limb lead mode that offers choices of the leads I, II, III, AVR, AVL, or AVF.
The standard mode is available if the chest lead is applied. The limb lead mode is
available when there is no chest lead applied. Loss of the chest electrode
changes the ECG - Lead Select menu to the limb lead mode if the left arm, left leg,
and right arm electrodes are intact.
Table 3: Two Operational Modes
!
With dual-lead transmitters, both modes work correctly with or
without the right leg electrode attached. However, for optimum
performance, the right leg electrode should always be used.
Connected Electrodes
(X)
91341
Valid Lead Vectors
CLALLRA
X X X X V1-6 and II (standard mode)
X X X III (standard mode)
X X X I (standard mode)
X X total lead failure
(standard mode)
X X X II (standard mode)
X X None (lead failure)
X X None (lead failure)
X None (lead failure)
X None (lead failure)
X None (lead failure)
X None (lead failure)
None (lead failure)
X X III (limb lead mode)
X X I (limb lead mode)
X X II (limb lead mode)
X X X I, II, III, AVR, AVL, AVF
(limb lead mode)
!
With dual-lead transmitters, if one of the leads fails, a lead fault
message will be displayed in the upper left corner of the
waveform zone. If there is no valid lead vector, the message
LEADS OFF is displayed and an alarm tone is generated.