User's Manual

SpO2 (91343 only)
071-0774-00 Rev. A 3-3
Figure 3-2: Display Zone — Full Screen
Figure 3-3: Display Zone — Split Screen
Current SpO
2
value (percent) and episodic time of reading. (Time is not
displayed when using continuous mode of operation.)
The bell indicates that alarms are enabled (equal sign turns to a small bell
when SpO
2
alarms are enabled).
SensorWatch bar:
Shaded area (waveform index, WFI) expands up proportionally to signal
strength; horizontal line is minimum signal level.
No shading (lowest waveform index) corresponds to no detected signal
strength or a faulty sensor.
Ensuring Accurate Monitoring
Each sensor requires site specific application procedures, and the following
general points will aid oximetry monitoring success.
Choose a site that provides proper alignment of the LEDs and receiving
photodetector.
Reduce light interference when monitoring under bright light by using a light
block over the sensor.
Select a site that has unrestricted blood flow and can remain as immobile as
possible to reduce or eliminate movement artifact.
Do not restrict blood flow when securing a sensor with tape.
Do not select a site near potential electrical interference (e.g., electrical
cords).
The SensorWatch bar should be above the minimum signal level.
E
C
G