User's Manual
EXT 1824G User’s Manual 1. Safety Information
EXT 1824G-UM-COM-EN-03 12
1.2. Notes for using the equipment
When using the equipment, take the following precautions. Otherwise, problems may occur and the equipment may
not function correctly.
System Diagnostic
• The Ecali1 software runs a system diagnosis.
• Run Ecali1 software after installing the system, at least once a year. If an error occurs, report the detailed error
information to DRTECH local dealer or distributor.
C
alibration
• To ensure optimal performance of the system, it is important to verify that the system is calibrated.
• You can process calibration with the calibration data CD (provided).
Before exposure
• Be sure to check the equipment on a daily basis and confirm that it works properly.
• Suddenly heating the room in clod area will cause condensation to form on the equipment. In this case, wait until
the condensation evaproates before performing an exposure. If the equipment is used while the condensation
forms, problems may occur in the quality of captured images. When an air-conditioner is used, be sure to
raise/lower the temperature gradually so that a significant change in temperature in the room and in the
equipment does not occur, to prevent condensation.
During exposure
• Do not use the selected frequency chanel (2.4GHz and 5GHz dual band) for other wireless devices.
Mutual interference may affect the image data transmission rate.
CAUTION
The owner is responsible for ensuring that the system diagnostic is
performed every year.
Do not try to use the system if the system diagnosis failed.
CAUTION
The owner is responsible for ensuring that the system calibration is
performed whenever the system installation is completed or the system is
repaired. Do not try to use the system if system calibration is not
performed.