7.2.2. Detector Power Save Management
Meaning
Mode Description
Normal Detector can be operated and take an exposure at any time.
Sleep Detector can be operated and take an exposure at any time within 2 seconds.
Deep Sleep Detector can not be operated. User can take an exposure only if Deep Sleep Mode is canceled.
Power Off Detector has been turned off. User can take an exposure after detector is rebooted.
Entry Condition for Power Save Mode
Mode Description
Normal -
Sleep Activates when the detector has not been used for a certain time(Default Value is 5 min.).
Deep Sleep
Time can be set to automatically go into Deep Sleep mode, after a certain period of time passed
under sleep mode(Default Value is 3 min.).
Power Off
Detector is turned off if power button on the detector is pushed
Escaping from Power Save Function
Mode Description
Normal -
Sleep
Sleep mode is turned off within 2 sec. when user enters into the examination window or gives
signal phisically such as push any button on the keyboard, shaking the mouse, etc.
Deep Sleep
Deep Sleep mode is turned off within 10 sec. when user gives signal phisically such as push any
button on the keyboard, shaking the mouse, etc.
Call the function for turning off Deep Sleep Mode at SDK.
Power Off
Reboot the detector with pressing power button on the detector.