EXT 1824G User’s Manual 9. Maintenance and Inspection
EXT 1824G-UM-COM-EN-03 62
9. Maintenance and Inspection
In order to ensure that the equipment is used safely and normally, be sure to inspect the equipment before use. If any
problem is found during the inspection and cannot be corrected, please contact your sales representative or local
DRTECH dealer.
Daily Inspection
WARNING
For safety reasons, be sure to turn OFF the power to each piece of
equipment before the following procedures. Otherwise, an electric shock
may result.
Cable
1) Ensure that cables are not damaged and cable jackets are not torn.
2) Ensure that the power cord plugs are securely connected to both the equipment AC inlet and the AC outlet.
Detector
1) Ensure that there are no loose screws or broken parts.
2) Ensure that there is no dust or foreign substance on the external connector.
3) Ensure that there are no broken parts or short-circuits in the power supply conector.
After turning on the power
Be sure to start ECali1 before performing the following inspection.
1) Perform test exposure.
Monthly Inspection
1) Ensure that there are no loose screws or broken parts.
2) Ensure that there is no dust or foreign substance on the external connector.
Yearly Inspection
1) Perform a Performance Test or Self-diagnosis using a phantom or resolution chart, etc.
Irregular Inspection
Calibration
1) Perform Calibration when exposure conditions have changed significantly.
For details, refer to the Setup Guide for ECali1.