Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) User’s Manual PEDRA-1417M Detector (Wired & Wireless) Before using this equipment, be sure to read this user’s manual thoroughly. Keep this user manual where it is easily accessible.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) To customers Thank you for purchasing a PEDRA detector. This User Manual explains how to use the detector, X-ray interface unit, and other peripheral equipment. Before using this product, be sure to read this manual thoroughly in order to utilize it more effectively.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Note on installation Request your sales representative or distributor to install this product. Notes on disposal of this product Disposal of this product in an unlawful manner may have a negative impact on human health and the environment.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Content Safety and Regulatory.......................................................................................................................................................7 Safety Notices ...................................................................................................................................................................7 Regulatory ....................
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.4 1. 3.5 4 Turn on the detector...................................................................................................................................... 36 Detector status list ................................................................................... 36 Turn on the Battery Charger....................................................................................
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Appendix.3 Specifications ............................................................................................................................................ 61 1. PEDRA-1417M Specifications .................................................................................................................... 61 2. PEDRA-17F Dimensional diagram .....................................
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Safety and Regulatory Safety Notices The following safety notices are used to emphasize certain safety instructions. Follow the safety instructions in this manual along with warnings and cautions symbols. Ignoring such warnings or cautions while handling the product may results in serious injury or accident.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 1 Safety Information 1.1 Safety precautions This product is designed and manufactured to ensure maximum safety of operation and to meet all the safety requirements applicable to electronic medical equipment. Follow these safeguards and properly use the equipment to prevent injury or damage to any equipment or data.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Do not connect a multiple portable socket-outlet or extension cord to the system. Doing so may result in a fire or electric shock. Warning Handling Never disassemble or modify the device. Doing so may result in a fire or electric shock.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Warning Installation and environment of use Before authorizing any person to operate the system, verify that the person has read and fully understood the User Manual. The owner should make certain that only properly trained and fully qualified personnel are authorized to operate the equipment. An authorized operators list should be maintained.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Do not spill liquid or chemicals onto the detector. In cases where the patient is injured, do not allow it to come in contact with blood or other body fluids. Doing so may result in a fire or electric shock. In such a situation, protect the device with a disposable covering as necessary. Turn OFF the power to each piece of equipment for safety when not being used.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 1.2 Symbols and detail of marking Symbol Description Direct Current Alternating Current This mark indicates that this equipment must be handled with care. Do not jolt or apply excessive load to the equipment. General warning sign Attention, consult accompanying documents Read and understand all instructions and warning labels in the product documentation before using the equipment.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) When using the device, take the following precautions. Otherwise, problems may occur and the device may not function correctly. Image Backup To avoid missing images which might result in the patient being exposed to additional doses of radiation, it is important to backup images by filming or using a CD or DVD option. This should be done as a routine operation for every patient.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) After using the detector and its peripheral device for examination, wipe the patient contact surfaces of the detector and peripheral device (attached handle included), by disinfectants such as IPA and ethanol to prevent infection. For details on how to sterilize, consult a specialist.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) The device is mainly for use in X-ray exposure rooms and hospital wards. To use it in other places, consult our sales representative or a distributor.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Detector & Power Supply for wireless communication) Do not expose this device to high temperatures and/or high humidity. Malfunctions may occur. When not in use, keep the detector, handle unit, and grid in a designated location or in a location where they are safe and cannot fall down.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2 Introduction PEDRA 1417M detector is a digital X-ray flat panel detector which has 35 ㎝ x 43 ㎝ imaging area and communicates with wired and wifi communication feature, Giga-bit Ethernet communication method through connecting a tether cable and wifi router.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.3 Detector Description The detector is designed to capture radiographic images. Captured images are transmitted to the viewer computer through a wired and wireless connection. A. POWER button: Controls the detector power on/off B. Status indicators a. Power LED: Indicates power on/off status (Blue). b. Network LED: Network Connection status of wired mode (Green) c.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) C. Tether port: Allows the detector to communicate with the Power Supply via tether cabling (Gigabit Ethernet)1000 BASE-T). D. Wifi Antenna E. Battery Connector (1) Battery connection parts (2) AP mode On/Off button: User can select AP mode or Wifi (or tether) mode.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.4 Power & Control Box (PEDRA-PCON) Description PEDRA-PCON is connected in between PEDRA detector and work station computer. This device supplies DC power to PEDRA detector, and receives orders from the computer. And it transmits digital images from PEDRA detector to the station. (1) External power socket: AC/DC Adopter [MPU-64], DC +15V / 4.2A INPUT (2) Tether cable port: Operation: a.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) (5) X-ray SYNC port: READY/EXPOSURE out port Operation: When an event occurs in normal open condition, RDY_OUT/EXP_OUT and RDY_COM/EXP_COM are closed. (6) Status lamps: a. PCON Power LED: Power on LED (Blue) b. Exposure LED: Generator X-ray exposure LED (Orange) / at DMS and Manual Mode activated c.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.5 Battery Charger (PEDRA-BC001) Description (1) Power input port: Supplies electric power by connecting a power adapter to the battery charger. (2) Power LED: Indicates the power status On / Off (3) Charging Status LED: Indicates charging status of a battery pack. (4) Battery insertion: Insertion part for battery charging 2.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.7 Components PEDRA consists of Detector, PEDRA-PCON, Generator SYNC, Tether Cables and some accessories.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Battery Charger Battery Adapter Battery AC Power Cable (230V & 120V) 1) Standard Part: Item Detector Power & Control Box Battery Charger Battery PEDRA CD Product Name PEDRA-1417M (scintillator: CsI (TI)): 3.0kg (GdOS): 3.20kg PEDRA-PCON: 0.50kg PEDRA-BACH001: 0.39kg PEDRA-BT001: 0.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2) Optional Part Item Product Name Hand Switch SYNC cable Hand Switch Handle Portable used Handle Manual Mode use USB Cable Caution The use of accessories and cables other than those specified, with the exception of PEDRA accessories and cables sold by Manufacturer.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.9 Grid (Recommended not included) Item Description SID 100 /130/150/180 cm Size 384mm x 460mm x 1.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3 Detector Installation PEDRA_1417M detector is properly adjusted at installation by a service engineer. If you encounter any problems during normal operation or daily inspections, consult your sales representative or distributor. 3.1 Preparations 1. Charge /Attach the battery pack Charge the battery on the day of examination or on the previous day.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Make sure that the battery pack is securely attached. The remaining battery charge can be checked on the monitor of viewer. You can use the detector while charging the battery by connecting tether interface.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) When the detector is not to be used for some time, remove the battery pack. Otherwise, over discharge may occur resulting in the shortened battery life. You can purchase an optional battery pack to replace an exhausted one.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.2 Connection 3.2.1 Tether Connection The PEDRA 1417M detector and Power box are connected with the tether cable. PEDRA 1417M detector has a tether interface cable. With the wiring connector, it is possible to switch from a wireless configuration to a wired configuration (see the figure).
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.2.2 Wireless Connection Wireless Connection mode is the wireless connection method that connecting the detector with wifi router as the center wirelessly. The PEDRA 1417M detector and Router are connected wirelessly. In this case, Router and Detector operate as AP (Access Point) and Station respectively. Router, Power box, and PC are connected with the LAN cable.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.2.3 Detector AP Connection Detector AP mode is the wireless connection method that connecting the PC (Workstation) with the detector as the center wirelessly. The PEDRA 1417M detector and PC are connected wirelessly. In this case, the detector and the PC operate as AP (Access Point) and STATION respectively.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.3 PEDRA-PCON Connection 1) To transmit image data and generator signal using a tether cable, connect the one end of the tether cable to the PEDRA-PCON and the other to the detector. 2) Connect the one end of the LAN cable to the PEDRA-PCON, and the other to the LAN Card Connector of workstation assigned for the Data Transfer. 3) 4pin SYNC cable connects hand switch and the PEDRA-PCON.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) a Lamp type [only for wired connection] b c PCON Power X-ray Exposure Detector Power Blue Orange Green O X X b. X-ray ready & Shot O O O c. O X O X X X Color a.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.4 Turn on the detector 1. Detector status list : Press the power button of the detector for 1.0 second, then the green LED is activated. And detector of Green LED is activated after 40 seconds of blinking every 0.5 seconds. Lamp type [Wired Connection] Lamp Color Active Blue Green Orange O X X b. Network Connection [Wired mode] X O O [blinking] c. X O O d.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3.5 Turn on the Battery Charger 1) Connector for Power adapter is positioned on left side of charger. 1 No.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 4 Detector Operation S/W Setting (PEDRA-DMS) Allow the detector for at least 30 minutes to warm up before use for maximum performance. Otherwise, it may result in faulty images until warm up has occurred. 4.1 Installation It needs to Copy ‘PedraCalib’ folder from the ‘PEDRA CD’ to the workstation computer. 1. “PedraCalib.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) (1) Path setting User can set work folder.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Calibration, image, log folders are automatically created in the set work folder. Calibration files are saved in the calibration folder. (dark, bright, BPM) The exposured image is saved in the image folder. Logs are saved in the log folder. (2) Detector information IP address & Port are displayed.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) - Sets effective area of an image. (3) Detector status check User can check dectector status (detector temp., wifi strength, battery level) - : User can check Detector status - : User can check Detector log - : User can check Detector status check data (4) Get image for calibration - : Get a dark image for offset calibration. - : Get a Bright image for gain calibration.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) ○ 1 Image list - saved image list from dark & bright files ○ 2 Selected Dark files - Dark files, selected from the dark files acquired by the user for calibration. User can select up to five for calibration. ○ 3 Selected Bright files - Bright files, selected from the bright files acquired by the user for calibration. User can select up to five for calibration.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) ○ 1 Load BPM & Save BPM - Load the BPM file that needs to be modified, or save the modified file. ○ 2 image tools - User can adjust zoom ○ 3 clear bad pixel - User can add or delete defect pixel Auto or manually for calibration. ○ 4 bad pixel location - User can check bad pixel location.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) (7) Log & Image view - User can check the PEDRACalib’s log. (Click once more to disappear.) - : User can check the dark or bright image. (Click once more to disappear.) - : Exit the PEDRACalib. 4.3 Calibration Procedure For calibration implementing, Manual or AED mode needs to be selected to acquire the secure image, dark and bright.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Step 0. Connect detector Input detector IP & port then Connect. Step 1. Selection of AED mode: For more detail information, refer to AED mode configuration, chapter, 5.2. Step 2. Click the “ ” button to start acquire Dark Image from the detector. The Dark Image will save as Dark_image resolution_bit_YYYYMMDD_HHMMSS.raw (The acquired dark frame will be generated in the selected folder.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Step 3. After Step3, Click the “ ” button The X-ray condition should be set or tested beforehand, Click the “Get Bright” (The acquired Bright frame “Bright_ image resolution_bit_bright value.raw” will be generated in the selected folder.) Bright value [LSB] is descrived on the “table 1” and then it will be acquired 5 different level of bright images. Step 4.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Step 5. Getting raw image: Select the AED or Manual mode and shoot the X-ray using the “Hand switch or shoot button on the X-ray Generator. The way of the image acquiring is referenced to AED or Manual mode operating, chapter 5.2 and 5.3.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 5 Operation 5.1 Wireless Standard Configuration and Operation Generally, the PEDRA detector is used in a wireless configuration as illustrated below: Wireless communication is established between the PEDRA detector and Router. The PEDRA detector is compliant with IEEE 802.11 a/b/g/n (2.4 ㎓ / 5 ㎓). The available frequency band may vary depending on local radio laws and system requirements.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 5.2 AED Mode Configuration and Operation in wired mode AED mode is available for acquiring images without any connection to the X-ray generator with an X-ray SYNC cable The way of image acquiring is as follows; 1) Selection of AED mode 2) Shoot the X-ray using hand switch of X-ray generator.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 5.3 Manual Mode Configuration and Operation in wired mode Hand switch is connected with the PEDRA-PCON to communicate with the PC via USB cable. Manual Mode is option to shoot the X-ray with Hand switch. Manual mode is a wired connection with tether cable and USB cable to communicate with PC via LAN cable.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 5.4 Detector AP Configuration and Operation Generally, the detector AP mode is used in a wireless configuration as illustrated below: Wireless communication is established between the PEDRA detector and PC. The PEDRA detector is compliant with IEEE 802.11 a/b/g/n (2.4 ㎓ / 5 ㎓). The available frequency band may vary depending on local radio laws and system requirements.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 6 Troubleshooting 6.1 General When you encounter problems or error messages while using this equipment, search the table below for the problem or error messages and try the solutions. If the problem persists, turn off the detector and consult your sales representative or a distributor. Please refer to the details of the problem or error messages.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 7 Maintenance In order to ensure that the device is used safely and normally, be sure to inspect the device before use. If any problem is found during the inspection and cannot be corrected, please contact your sales representative or a distributor. Daily inspection For safety reasons, be sure to turn OFF the power to each piece of device before attempting the following.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Function test Item Power consumption Period Daily Description Confirm that the power operation of detector is normal. Check the monitoring in order to minimize the characteristic Temperature Daily changes of Flat panel caused by external temperature changes.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 8 Information 8.1 Service information Product lifetime The estimated product lifetime may be up to five (3) years under appropriate regular inspection and maintenance.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 8.2 Warranty Radisen warrants that this product will be free from defects in materials and workmanship for a period of two (2) years from the date of delivery.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 9 Revision History Revision 0.0 Date 2020.07.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex.1 Generator Interface 1. X-ray SYNC Connect the X-ray SYNC cable to the PEDRA-PCON, 6pin SYNC and the X-ray generator. 2. Tether Cable Pin assignment Tether Cable is female connector to connect between Detector and PEDRA-PCON.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Appendix.2 Detector IP setting 1. Default IP address wired or wireless configuration if the connection is unstable or setting value information is not clear when you intend to change it. Wire Mode Wireless HUB Mode Wireless Access point Mode PC 192.168.0.55 - - Subnet Mask 255.255.255.0 255.255.255.0 - PEDRA 192.168.0.11 192.168.0.11 11.11.11.1 2.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Appendix.3 Specifications 1. PEDRA-1417M Specifications Model & Parameters PEDRA-1417MG PEDRA-1417MC GD2O2S:Tb CsI:Tl (Direct) Scintillator Sensor Image Pixel size 140um, a-Si TFT active matrix with PIN diode Pixel Matrix 2500 x 3052 Effective Area 350.0 x 427.0 mm perfor- Spatial Resolution mance 3.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3. PEDRA-PCON Item AC/DC Adapter Description Input: 100-240VAC~, 47~63Hz, 1.62-0.72 A Output: DC+15V /4.2A (max) Gigabit Ethernet Ports –1EA Cabling Ports USB port - 1EA X-ray SYNC - 1EA HAND switch – 1EA Dimensions (W ×H ×D) 200 ㎜×115 ㎜×36.2 ㎜ Weight 0.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 4. Battery Pack This dedicated battery pack is necessary for PEDRA wireless detector. Item Description Model PEDRA-BT001 Type Lithium Polymer Rated Power Supply Output: DC +7.58V Capacity 4000㎃h Number of Cell 3S1P (3 Series 1 Parallel) Life Approx. 500 times (charge/discharge 1cycle) Dimension (W ×H ×D) 219 ㎜×166.5 ㎜×8.0 ㎜ Weight 0.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 5. Battery Charger Item Description Model PEDRA-BACH001 Charging 1 battery packs Charging Time 2.5 hours Input: 100-240V~, 50/60Hz, 1.5 A Rated Power Supply Output: DC 15V, 4.0A Dimension (W ×H ×D) 250.0 ㎜×120.0 ㎜×161.5 ㎜ Weight 0.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Appendix.4 Label and Symbols PEDRA detector and other components have labels and markings on them. Their contents and locations are indicated below. 1.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 2.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 3. Battery Charger 4.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 5.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex . 5 Regulatory Information 1.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) European Union MDD (93/42/EEC) Medical Device Directive EN ISO 13485:2012+AC:2012 Medical devices –Quality Management systems –Requirements for regulatory purposes EN 60601-1: 2006 Medical electrical equipment-Part1: General requirements for safety IEC 60601-1-2: 2007(ed.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex .6 Serial Number (S/N) composition The serial numbers for each product and accessory are composed as follows; Product M 0 Model Code Rev (Fixed) G F A Sensor Type Secret Type Code Code A H 1 Year Month 0 0 1 Place Code Serial number Accessory PC/CH/BT 1 M 0 A H Model Code Series Components Rev Year Month 0 0 1 Serial number 1.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) 9. Composition of serial number for each item Item Serial No.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex .7 Guidance and Manufacturer’s Declaration for EMC This device has been tested for EMI/EMC compliance, but interference can still occur in an electromagnetically noisy location. Attempt to maintain a suitable distance between electrical devices to prevent malfunction.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Electromagnetic Immunity The PEDRA is intended for use in the electromagnetic environment specified below. The customer or user of this device should assure that it is used in such an environment.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) turbances In- Stand by 240 V,50 Hz duced by 240 V, 60 Hz RF fields 6 V in ISM bands Between 0.15 MHz SIP/SOP and 80 MHz 80% AM at 1 kHz Power Frequency Magnetic Field Immunity Operating EN 61000-4-8:2010 120 V, 60 Hz Enclosure IEC 61000-4-8:2009 Stand by 220 V, 60 Hz 230 V,50 Hz 30 A/m 50 Hz & 60 Hz 0 % UT: 0.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex. 8 X-ray condition (Referenced) X-ray Energy Range: 40kVp ~ 150kVp Below exposure conditions were set by CPI generator, CMP200 DR for customer to reference. Large No.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex. 9 Wireless Communication Specification Wireless LAN Specification Item Description Wireless Standard IEEE802.11ac/a/b/g/n Frequency Range 2.412GHz~2.484GHz, 5.150MHz~5.850MHz 802.11b: 11Mbps / 802.11a/g: 54Mbps / 802.11n: MCS0~7/ 802.11ac: Data Rate MCS0~9 Bluetooth: 1Mbps, 2Mbps and up to 3Mbps EDR WiFi : 802.11b: DSSS (DBPSK, DQPSK, CCK) 802.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Network IP Address : 192.168.0.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) Annex. 11 FCC-Related information and explanation Installation Procedure FCC Caution: Any changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate this equipment.
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1) IMPORTANT NOTE: FCC Radiation Exposure Statement: This device meets FCC limits for exposure to radio waves. It is designed not to exceed the limits for exposure to radio waves (radio frequency electromagnetic field) adopted by the Federal Communications Commission. This equipment should be installed and operated with minimum distance 0mm between the radiator & your body.