User's Manual

“More than just Digital” Page 26 of 81
Doc. Name: PEDRA Users Manual/Doc. No.: PEDRA-1417M-UM
Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1)
2) Optional Part
Item Product Name
Hand Switch SYNC cable Hand Switch
Handle Portable used Handle
Manual Mode use USB Cable
Caution
The use of accessories and cables other than those specified, with the exception of PEDRA accessories and
cables sold by Manufacturer. as replacement parts for internal components, may result in increased radiation
emissions or decreased stability of the equipment. Accessory device connected to the analog and digital
interfaces must be certified according to the respective IEC standards. All combinations of device must be
compliance with IEC 60601-1 system requirements. Any person who connects additional device to the signal
input or signal output ports configures a medical system, and is therefore responsible for ensuring that the
system complies with the requirements of the system standard IEC 60601-1. If in doubt, consultmanufacturer
technical support representative.
2.8 Workstation (Recommended)
Item Specification
Operating System Windows 8, 10, 64 bit SP1 (Professional Edition or higher)
CPU Intel Core i5 2600 or higher (or compatible CPU)
Memory 4GB or higher
Hard Disk 1TB or higher
LAN Card
Gigabit (Detector only)
Intel? PRO 1000 Series
(Gigabit LAN Card for network interface)
Min. Requirements: 1Gbps, Jumbo Frames: 9K
Receive Descriptors: 2K (higher than 1024)
This is not dedicated to DICOM
Monitor
1024 X 768 or higher
Optical Disc Drive
CD or DVD R/W