User's Manual
“More than just Digital” Page 42 of 81
Doc. Name: PEDRA User’s Manual/Doc. No.: PEDRA-1417M-UM
Initiate date: 2020.07.25 /Revision date: 2020.10.15 (Rev. 1)
- Sets effective area of an image.
(3) Detector status check
User can check dectector status (detector temp., wifi strength, battery level)
- : User can check Detector status
- : User can check Detector log
- : User can check Detector status check data
(4) Get image for calibration
- : Get a dark image for offset calibration.
- : Get a Bright image for gain calibration. When you click the button detector
ready for get X-ray image
- : Get a raw image for image check.
- : Call the last raw image.
- : Cancel calibration.
(5) RUN Calibration
- A calibration file (BPM) is created using the saved dark and bright files.
- if user need to add bad pixel or line, user can clear.
- : user can make calibration.