3.2.2 Software Sync Mode
The detector communicates with an X-ray generator via the DROC workstation via a wired or wireless
network connection. Compared with the external sync mode, the software sync mode will cause a
delayed response.
The sequence chart of software sync mode for a complete cycle of image acquisition is shown in figure
below.
Hand Switch 2
nd
Gear
X-ray ON
pressed
Exposure Time
released
SOFT_SYNC_EXP_REQ
SOFT_SYNC_EXP_EN
Figure 10 Sequence chart
A typical configuration for software sync mode of a DR system integrated with CareView 750Cw is
shown in the figure on the following page.