Operation Manual
Page 22 of 57
RD-OI-20 A/3
Note:
1) In manual sync mode, the image will be light and have an uneven shade acquired in the forbidding
exposure time window.
2) If no goal image appears, first verify the exposure dose to see whether it is too low.Then check the
tube’s field of view to verify that it completely covers the plate.
The sequence chart of manual sync mode for a complete cycle of image acquisition is shown in figure
below.
Hand Switch 2
nd
Gear
X-ray ON
pressed
Exposure Time
released
Figure 13 Sequence chart
A typical configuration for manual sync mode of a DR system integrated with the CareView 750Cw is
shown in the figure below.
X-ray Tube
X-ray Generator
Wireless AP
DROC Workstation
Shielded X-ray Room
Gigabit LAN Cable
IEEE 802.11n
CareView 750Cw
Control Panel
Figure 14 Manual sync mode
A typical configuration for manual sync mode of a DR system integrated with the CareView 750C is
shown in the figure on the following page.