Operation Manual
Page 54 of 57
RD-OI-20 A/3
WARNING: Use of this equipment X-ray flat panel detectors adjacent to or stacked with other
equipment should be avoided because it could result in improper operation.”
When the AC input voltage is interrupted, the X-ray flat panel detectors will power off and if the
power supply restored, it should be recovered by operator manually, this degradation could be
accepted because it will not lead to unacceptable risks and it will not result in the loss of basic
safety or essential performance
EMI Compliance Table
Table 23 Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class A
Professional healthcare facility environment
Harmonic
distortion
IEC 61000-3-2
Class A
Professional healthcare facility environment
Voltage fluctuations
and flicker
IEC 61000-3-3
Compliance
Professional healthcare facility environment
EMS Compliance Table
Table 24 Enclosure Port
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare facility environment
Electrostatic
Discharge
IEC 61000-4-2
±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM
field
IEC 61000-4-3
3V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
Refer to table 3
Rated power
frequency magnetic
fields
IEC 61000-4-8
30A/m
50Hz or 60Hz
Table 25 – Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Professional healthcare facility environment
385
380-390
Pulse modulation 18Hz, 27V/m