RD-OI-20 A/3
Table 1 Typical patient doses
Body part
Average adult
patient size
(with Anti-scatter
grid, 40 lp/cm 10:1)
SID(m)
Dose
(μGy)
kV
mAs
Chest, PA
Height: 170cm
Weight: 65kg
112
4
1.5
176
L-spine, AP
75
40
1
1770
Pelvis, AP
65
32
1
910
Hand, AP
50
2.5
1
38
Knee, AP
65
5
1
150
Foot, AP
50
2.5
1
38
Note:
PA: Postero-Anterior position; AP: Antero-Posterior position
Dosage amount is typical for average-sized adult patients.
1.4 Configuration
CareView 750Cw detector supports both wired and wireless configurations, while CareView 750C
detector only supports wired configuration, as illustrated below.
Figure 1 Wired configuration
X-ray System
Generator
DROC Workstation
Power Box
Gigabit LAN cable
Interface cable
CareView 750C
CareView 750Cw